Principle and formula of magnetic permeability test method

The magnetic permeability μ is equal to the ratio of the differential of the magnetic induction B in the magnetic medium to the differential of the magnetic field strength H, ie μ=dB / dH

Generally used is the relative magnetic permeability μr of the magnetic medium, which is defined as the ratio of the magnetic permeability μ to the vacuum magnetic permeability μ0, that is, μr=μ/μ0

The relationship between the relative magnetic permeability μr and the magnetic susceptibility χ is: μr=1+χ

The magnetic permeability μ, the relative magnetic permeability μr and the magnetic susceptibility χ are physical quantities describing the magnetic properties of the magnetic medium.

For the paramagnetic mass μr>1; for the antimagnetic mass μr<1, but the μr of both are almost the same as 1 . In most cases, the relative permeability of the conductor is equal to 1. In ferromagnetic properties, the relationship between B and H is a nonlinear hysteresis loop, and μr is not a constant, and is related to H, and its value is much larger than 1.

For example, if the relative magnetic permeability of air (non-magnetic material) is 1, the relative magnetic permeability of the ferrite is 10,000, that is, when compared, the magnetic flux density through the magnetic material is 10,000 times. The cast iron is 200~400; the silicon steel sheet is 7000~10000; the nickel zinc ferrite is 10~1000.

Method principle

Speaking of the measurement of magnetic permeability μ, it seems very simple, just a few coils around the material sample ring, measuring its inductance, find a formula is complete. In fact, for the same sample loop, different instruments, different voltages, different voltages or different frequencies may be used to measure the magnetic permeability that is far different. The reason for the great difference in test results is that not every tester has the energy to make it clear. This paper mainly discusses the influence of different test parameters and calculation formulas on magnetic permeability measurement.

2.1 Influence of calculation formula

As we all know, the method of measuring the magnetic permeability μ is generally to measure the inductance L around the N 匝 coil on the sample loop, because the expression that can be derived from L is:

L=μ0 μN^2A/l (1)

Therefore, the formula for calculating the permeability from (1) is:

μ=Ll/μ0N^2A (2)

Where: l is the magnetic path length of the core, and A is the cross-sectional area of ​​the core.

For a toroidal core with a rectangular cross section, if its average magnetic path length l = π (D + d) / 2 is taken as the magnetic path length l of the core, the sectional area A = h (Dd) / 2, 00=4π×10-7 are substituted into (2):

μ=L(D+d)*10/4Nh(Dd) (3)

Where D is the outer diameter of the ring, d is the inner diameter, and h is the height of the ring, as shown in FIG. Substituting the inner diameter d=D-2a of the ring into equation (3):

μ=L(Da)*10/4Nha (4)

Where: a is the wall thickness of the ring.

For a toroidal core with a smaller inner diameter, the inner diameter is not as easy to measure as the wall thickness, so it is convenient to use the type (4). Equations (4) and (3) are equivalent, and their origin is to take the average magnetic path length of the ring as the magnetic path length of the core. The magnetic permeability calculated from them is called the ring magnetic permeability of the material. Some people say that the magnetic permeability measured by a ring sample is called ring magnetic permeability. This is not true. In fact, the ring permeability is higher than the actual magnetic permeability of the material, and the thicker the wall of the sample ring, the greater the error.

For the sample ring, the magnetization field is non-uniform in the radial direction under the same ampoule magnetomotive force excitation. The closer to the outer side of the ring wall, the weaker the magnetic field. Under the condition that the magnetic permeability μ is constant throughout the sample ring, the closer to the outer side of the annular wall, the lower the magnetic flux density B of the ring. In order to eliminate the influence of this uneven magnetization on the measurement, we consider the sample ring to be composed of an infinite number of thin-walled rings with a radius r and an infinite thickness of dr. According to formula (1), the inductance dL produced by each thin-walled ring can be written as:

(5)

From (5), the integral r is integrated from the inner radius r1 to the outer radius r2, and the inductance L generated by the entire sample ring is obtained:

(6)

The exact formula for calculating the magnetic permeability derived from equation (6) is:

(7)

In order to facilitate practical application, (7) can be converted into;

(8)

In the above formula: D is the outer diameter of the sample ring, and d is the inner diameter. Converting the natural logarithm to a common logarithm, (8) is transformed into:

(9)

If the sample ring is composed of the same material, the magnetic permeability calculated by the formula (7), (8) or (9) is the true magnetic permeability μ of the material. It is slightly lower than its ring permeability.

2.2 Test the influence of the number of turns of the coil N

Since the inductance L is proportional to the number of turns N2, it is reasonable to say that the magnetic permeability μ calculated by the formula (9) should not be related to the number of turns N, but it is often related.

Regarding the measurement of the magnetic permeability of materials, the test frequency generally used is not high, and is often tested at a frequency of 1 kHz or 10 kHz. The test signal is generally a sinusoidal signal, because the frequency is not high, the resistance part of the sample surrounding the coil impedance is negligible, and the winding coil is regarded as a pure inductance L connected to the measuring instrument. The test equivalent circuit is shown in the figure. The effective value of the voltage generated by the instrument signal source is U, and Ri is the output impedance of the signal source. The expression of magnetizing current is easily written by Figure 3:

(10)

In the above formula, ω is the angular frequency of the instrument signal source, and L is the inductance of the sample surrounding coil.

L=μ0μN2Ae /le (11)

In (11), Ae is the effective sectional area of ​​the core, and le is the effective magnetic path length of the core. If the Ae and le of the ring core are substituted, the equation (11) becomes the same as the result of the equation (6).

The effective magnetic field strength peak Hm produced by the test current is:

(12)

Substituting (10) and (11) into (12) gives:

(13)

It can be seen from equation (13) that when (ωμ0μAe) 2N4 is much smaller than le2Ri2, the equation (13) can be approximated as:

(14)

The above formula tells us that when the number of turns of the test coil is small, the strength of the test magnetic field is proportional to the number of turns. As the number of turns increases, when (ωμ0μAe) 2N4 is much larger than le2Ri2, the equation (13) can be approximated as:

(15)

It can be seen from equation (15) that when the number of turns of the test coil is too large, the strength of the test magnetic field is inversely proportional to the number of turns.

From the above analysis, when measuring the magnetic permeability, the magnetization field strength in the sample loop is related to the number of turns of the test coil, and when the number of turns is a certain value, the magnetic field strength reaches the strongest value. The magnetic permeability of the material is closely related to the magnetization field strength, so the measurement of the magnetic permeability is related to the number of turns of the test coil. The effect of the number of turns on the permeability test is discussed in detail in conjunction with the figure.

2.2.1 When the test voltage U is low

As mentioned earlier, for high-end instruments, such as the Agilent 4284A precision LCR tester, its test voltage can be adjusted so low that the test magnetic field strength is the strongest with the change in the number of turns, and there is still no magnetic permeability. Start area. At this time, the initial permeability μi of the material is always measured, which is independent of the number of turns of the test coil. With the same instrument, if the test voltage is adjusted to a high level, it can no longer be guaranteed that the magnetic permeability measured by different turns is the initial permeability, and the measured magnetic permeability will be compared with the test coil turns. It’s related.

2.2.2 Test voltage U can not be adjusted

Most simple instruments that measure inductance cannot be flexibly adjusted in terms of test voltage and frequency. For example, the 2810 LCR bridge has a test frequency of 100 Hz or 1 kHz and a test voltage of less than 0.3V.

formula

Energy density of magnetic field = B^2/2μ

In the International System of Units (SI), the relative magnetic permeability μr is a dimensionless pure number, and the unit of magnetic permeability μ is Henry/meter (H/m).

Commonly used vacuum permeability

Features

The measurement of magnetic permeability is indirect measurement, the inductance of the winding coil on the core is measured, and the magnetic permeability of the core material is calculated by the formula. Therefore, the magnetic permeability test instrument is the inductance tester. It is emphasized here that some simple inductive test instruments cannot be adjusted in test frequency and the test voltage cannot be adjusted. For example, some bridges have a test frequency of 100 Hz or 1 kHz and a test voltage of 0.3 V. This 0.3 V is not the voltage across the inductor, but the voltage generated by the signal generator. As for the voltage across the coil under test is an unknown number. If you use a high-end instrument to measure inductance, such as the Agilent 4284A precision LCR tester, not only the test frequency is adjustable, but also the voltage and magnetizing current across the measured inductor coil are adjustable. Understanding these features of the test instrument can be a great help for the correct measurement of magnetic permeability.

Beijing Heng Aode Instrument Co., Ltd. specializes in stainless steel water collector, tank bottom weld vacuum test box, reading instrument, eight-stage air microbial sampler, relay comprehensive tester, dual-wavelength scanner, coating thickness gauge, soil Crusher, tempered glass surface flatness tester, sound sensor, portable electric water level gauge, network mouth flow meter, corrosion rate meter, portable scratch tester, freezing point tester, water quality tester, online ammonia tester, coating Thickness gauge, coating thickness gauge, soil pulverizer, digital thermometer, gas sampling pump, ceramic impact tester, automatic crystallization point tester, drug freezing point tester, reed switch tester, constant temperature water bath , gasoline root transfer, gas sampling pump, tempered glass tester, water quality tester, PM2.5 tester, inhalable particulate matter detector, high frequency heat sealing machine, strain control triaxial instrument, milk somatic cell detector, helium gas concentration Detector, soil moisture conductivity tester, field strength meter, collection box, color ratio tester, capillary water absorption time measuring instrument, redox potentiometer vibrometer, Carbon oxide carbon dioxide detector, CO2 analyzer, oscillopolarograph, slime content tester, automotive starter power supply, automatic potentiometric titrator, portable thermometer, zirconia analyzer, reed switch tester, precision conductivity Instrument, TOC water quality analyzer, microcomputer plasticity measuring instrument, wind direction station, automatic spotting instrument, soil oxidation reduction potentiometer, digital thermometer, portable total phosphorus tester, corrosion rate meter, safety valve grinding tool, company adhering to " "Customer first, forge ahead" business philosophy, adhere to the "customer first" principle to provide quality services to our customers. Welcome patrons!