Gas chromatography is a powerful analytical technique that involves vaporizing the sample in the injection port and carrying it through a column using a carrier gas. As the sample components travel through the column, they are separated based on their differing affinities for the stationary phase. Once separated, the compounds are sequentially detected, allowing the instrument to generate signals that provide information about each component.
In addition to qualitative and quantitative analysis, gas chromatographs can also be used to determine various physical and chemical properties of the sample, such as partition coefficients, activity coefficients, molecular weight, and specific surface area. These parameters are essential for understanding the behavior of substances during separation and their interactions with the stationary phase.
When operating a gas chromatograph, three key factors should be carefully considered:
1. **Gas Purity**
The purity of the gases used in a gas chromatograph must meet strict standards—typically above 99.99%. However, many operators overlook the fact that different detectors may have varying requirements for gas purity. Using low-purity gases can lead to unstable baselines, excessive noise, or even erratic peaks. For example, if hydrogen with only 98% purity is used in a detector operating at high sensitivity (e.g., 104 MΩ), the baseline may become highly unstable, resulting in continuous fluctuations. Similarly, impurities in the carrier gas, such as trace oxygen, can shorten the lifespan of the capillary column.
2. **Gas Flow Ratios**
When using a flame ionization detector (FID), the correct ratio of nitrogen, hydrogen, and air is crucial for stable and sensitive operation. The ideal ratio for air to hydrogen is typically between 6:1 and 8:1, while the nitrogen to hydrogen ratio should be around 1:0.85 to 1. Ensuring an excess of air helps achieve complete combustion of hydrogen, which is necessary for maintaining a stable flame. Proper gas ratios not only enhance detection stability but also ensure accurate quantification and reliable correction factors.
3. **Environmental Conditions**
Gas chromatographs do not require extreme temperature control, but they generally perform best within a room temperature range of 5°C to 35°C. Humidity levels should be maintained between 20% and 85%. In high-humidity environments, some instruments may experience reduced insulation performance, leading to lower response values, especially in high-sensitivity applications. It’s important for users to monitor environmental conditions and take appropriate measures, such as using dehumidifiers or placing the instrument in a controlled environment, to maintain optimal performance.
By paying attention to these factors, users can maximize the accuracy, reliability, and longevity of their gas chromatographic systems.